Sensors
Frequency band: 2 to 100 kHz
Detection range: 1.6 to >393' (0.5 to >120 m)
Field of view (FOV): 63° ±5°
Nominal frame rate: 25 FPS
Built-in Digital Camera (Visible Light)
Field of view (FOV): 63° ±5°
Focus: Fixed lens
Zoom: 3x digital zoom
Resolution: 5.0 MP
Display
Size: 7" (17.7 cm) LCD with backlight, sunlight readable
Resolution: 1280 x 800 (1,024,000 pixels)
Touchscreen: Capacitive
Acoustic image: SoundMap™ image
Image Storage
Storage capacity: 32 GB (>8000 pictures / >1600 videos)
Image format: Blended visual and SoundMap™ (.JPG or .PNG)
Video format: Blended visual and SoundMap™ (.MP4)
Video length: Up to 5 minutes
Digital export: USB-C for data transfer
Acoustic measurements
Measurement range (typical):
12.1 to 114.6 dB SPL (±1 dB SPL 2 kHz)
4.4 to 101.2 dB SPL (±2 dB SPL 19 kHz)
12.8 to 119.2 dB SPL (±1 dB SPL 35 kHz)
19.8 to 116.1 dB SPL (±3 dB SPL 52 kHz)
41.4 to 129.0 dB SPL (± 1 dB SPL 80 kHz)
54.4 to 135.5 dB SPL (± 1 dB SPL 100 kHz)
Classification and Quantification
Leak detection: Leak size scale, leak rate, and leak cost estimation (1 to 10)
PDQ Mode™: Partial discharge type classification
MecQ™: Mechanical anomaly detection
Auto Max/Min dB Gain: Auto or manual, user selectable
Frequency band selection: User selectable through user-made presets or manual entry
Software
User interface: Intuitive
Acoustics: Show dB scale: On/Off
Min/max dB: Manual/Auto
Profiles: Preset capture profiles
Multiple sources: Show single or multiple sources
Palette colors: Grayscale, ironbow, blue-red
Overlay transparency: Yes (1 to 100%)
Grayscale mode: Yes
Centerpoint markers: On/Off
Trend graphs: Frequency and dB scale
Spot markers: dB level reading at the center point of image
Leak values: Leak costs, leak volume, LeakQTM scale (1 to 10)
Master admin menu: Yes
Annotation: Asset name; Asset ID; Asset type; Leak type; Gas type; Pressure
Photo notes: Up to 4 photo notes for reference
Asset inspection status: 'As left', 'As found', 'Undetermined'
Asset ID: Text input and QR code-based identification
Actions annotation: Action requirements, priority level, action notes
Other: Operating conditions, weather notes
Fluke Premium Care
Premium care standard (sold separately)
Self-Diagnostic Notification
Array-health test to identify when microphone array needs attention
Tripod Holder/Connector
1/4"-20
Supported Languages
Dutch, English, Finnish, French, German, Italian, Japanese, Korean, Polish, Portuguese, Simplified Chinese, Spanish, Swedish, Traditional Chinese
Safety
IEC 61010-1: Pollution degree 2
Ingress protection (IP): IP40
Compatibility (EMC): IEC 61326-1: Portable Electromagnetic Environment, CISPR 11: Group 1, Class A
Korea (KCC): Class A Equipment (Industrial Broadcasting and Communication)
USA (FCC): 47 CFR 15 subpart B. Exempt device (15.103)
Environmental Conditions
Operating temperature: 14 to 113°F (-10 to 45°C)
Storage temperature: -4 to 158°F (-20 to 70°C) without batteries installed
Relative humidity: 10 to 95% non-condensing
Power
Battery type: Rechargeable and field replaceable Li-ion battery
Battery life: 6 hours
Battery charging time: 4 hours
Battery charging system: Internal (USB-C)
Dimensions
7.3 x 12.7 x 2.7" (186 x 322 x 68 mm)
Weight
4.4 lbs (2 kg) battery included