Fluke 4180 Infrared Temperature Calibrator, 5 to 248°F

Part Number : 4180-156

Brand :
Fluke

Increasing your IR temperature measurement accuracy has never been easier with this precision infrared calibrator. Intuitive and easy-to-use, the calibrator provides accuracy and traceability for infrared thermometers and thermal imagers and reaches temperatures from 5 to 248°F (-15 to 120°C).

View all

KWD 3,676.910


Quantity



Description
Specifications
Downloads
Related products

Description

The Fluke 4180-156 Offers Coming from one of the world’s most trusted temperature calibration laboratories, this precision infrared calibrator is incomparable, and the perfect solution for any infrared thermometer or thermal imager within its temperature range. Reliable with accuracies as good as ±32.6°F (0.35 °C), the calibrator can meet its specifications without additional emissivity-related corrections, leading to legitimate test uncertainty ratios (TUR) as good as 4:1. Features: Large target Dimensions of 152mm (6") captures "peripheral vision" of infrared thermometers and minimizes Dimensions of source effect errors Compensates for errors caused by individual infrared thermometer emissivity settings Sample calibration procedures for Fluke and Raytek thermometers 3are built right in Accuracy as good as ±0.35°C Test Uncertainty Ratios (TUR) as good as 4:1 Excellent stability and uniformity Reaches temperature setpoint quickly Intuitive, easy-to-read display that indicates when temperature is stable

Downloads

Fluke_Temperature_Datasheet_31.pdf

pdf

Specifications

Temperature Range (at 23°C ambient, 0.95 emissivity)
59 to 248° (15 to 120°C)
Power
115 V ac (±10%), 50/60 Hz, 250 W 230 V ac (±10%), 50/60 Hz, 250 W
Display Accuracy
±0.40°C at -15°C, ±0.40°C at 0°C, ±0.50⪚C at 50°C, ±0.50°C at 100°C, ±0.55°C at 120⪚C
Stablility
± 0.10⪚C at -15°C, ± 0.05°C at 0°C, ± 0.10°C at 120⪚C
Uniformity (5' (12.7 cm) diameter of center of target)
±0.15°C at -15°C, ±0.10°C at 0°C, ±0.25°C at 120°C
Uniformity (2' (5.08 cm) diameter of center of target)
±0.10°C at -15°C, ±0.10°C at 0°C, ±0.2°C at 120°C
Heating Time
15 min: 5 to 248°F (-15 to 120°C) 14 min: 73.4 to 248° (23 to 120°C)
Cooling Time
15 min: 248 to 73.4°F (120 to 23°C) 20 min: 73.4 to 5°F (23 to -15°C)
Stabilization Time
10 minutes
Nominal Emissivity
0.95
Thermometer Emissivity Compensation Range
0.9 to 1.0
Ambient temperature
5 to 35°C (41 to 95 °F)
Ambient Relative Humidity
maximum 80% for temperature <31°C, decreasing linearly to 50% at 40°C
Fuses
115 V: T 6.3 A, 250 V 230 V: T 3.15 A, 250 V
Altitude
<6561' (2000 m)
Target Diameter
6" (152.4 mm)
Computer Interface
RS-232
Dimensions
356 x 241 x 216 mm (14 x 9.5 x 8.5 in)
Weight
20 lbs (9.1 kg)
Safety
IEC 61010-1: Overvoltage Category II, Pollution Degree 2 IEC 61010-2-010
Industrial Spare Logo

Industrial Spare
Typically replies within an hour

Industrial Spare
Hi there,

How can I help you?
+96566651231
×
Chat with Us