Wika Cable Resistance Thermometer, TR41, -50 ... +250 °C, Tubular design

TR41.webp

Part Number : TR41

Brand : WIKA

Cable resistance thermometers are particularly suited to those applications in which the metallic probe tip is mounted directly into bored holes (e.g. in machine components) or directly into the process, and for any application with no chemically aggressive media or abrasion.

Refer to the datasheet or contact us for precise details.

View all
Quantity In Stock



Description
Specifications
Downloads
Related products

Description

Cable resistance thermometers are particularly suited to those applications in which the metallic probe tip is mounted directly into bored holes (e.g. in machine components) or directly into the process, and for any application with no chemically aggressive media or abrasion. A large number of different explosion protection approvals are availabe for the TR41. For mounting into a thermowell, a spring-loaded compression fitting should be provided, since only this can press the measuring tip into the bottom of the thermowell. Otherwise a potentially critical force could be exerted on the measuring tip. In the standard version the cable probes are manufactured without process connections. Fastening elements such as threaded connections, compression fittings, etc. are possible as options.

Applications

For direct installation into the process Machine building Motors Bearings

Special features

Sensor ranges from -50 ... +250 °C [-58 ... +482 °F] For insertion or screw-in with optional process connection Cable from PTFE, PFA, silicone and other cable sheath materials Versions with/without connector or connection housing (option) Explosion-protected versions are available for many approval types

Downloads

TR41 Data sheet TE 60.41

41

TR41 Operating Instruction

TR41 Additional operating instruction #1

TR41 Additional operating instruction #2

TR41 Additional operating instruction #3

Specifications

Specifications according to data sheet
TE 60.41
Design
With lead wires
Sensor
1xPt100, 3-wire
Accuracy class of the sensor
Class B, IEC 60751
Temperature range of selected class
-50...+250°C
Lead wire insulation
Silicon jacket, shielded, application range -60...+200 °C
Cable Length
2000 mm
Cable ending
Stripped leads
Probe material
Stainless steel 1.4571
Mechanical requirements
Standard probe tip (max. 6 g, peak to peak)
Stripped length
50 mm
Electrical approval
Without, ATEX
Ignition protection type
Without, Ex i
Probe diameter
6 mm, 3 mm
A-dimension
50 mm
Industrial Spare Logo

Industrial Spare
Typically replies within an hour

Industrial Spare
Hi there,

How can I help you?
+96566651231
×
Chat with Us