Fluke 718-30US Pressure Calibrator with Pressure Vacuum Pump, -12 to 30 PSI

718 30G.webp

Part Number : 718 30G

Brand : Fluke

Take pressure source and milliamp measurements to calibrate and maintain almost any pressure device.

View all

$ 3683.99


Quantity In Stock



Description
Specifications
Downloads
Related products

Description

Fluke 718-30US Offfers The Fluke 718 Pressure Calibrator provides a total pressure calibration solution for transmitters, gauges and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field. This meter provides a total pressure calibration solution for transmitters, gauges and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field. The 718 presents cost-effective solutions for highly accurate pressure measurements over a wide range of temperatures. Connect it with any of the Fluke pressure modules to address a wide variety of gauge, differential, absolute, vacuum and dual (compound) pressure applications. The Fluke 718 Pressure Calibrator provides a total pressure calibration solution for transmitters, gauges and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field. Features Error % calculation for quicker pass/fail decision making in the field Best in class mA accuracy of 0.015 % for better measurement confidence and workload coverage Min/Max/Hold to capture changing measurements Switch test feature for quicker and more accurate pressure switch testing New 1 and 300 PSI pressure ranges mean fewer extra tools required 1, 30, 100 & 300 PSI ranges now available Applications Pressure Instrument Calibration Pressure Instrument Maintenance Petro-Chemical Oil and Gas Pharmaceutical Water and Wastewater Food and Beverage Power Generation .fluke-pressure-calib-select td{border:2px solid #fff;text-align:center}.fluke-pressure-calib-select thead tr td{background:#95abc3;font-weight:bold}.fluke-pressure-calib-select thead tr td:first-child{background:#4a7696}.fluke-pressure-calib-select thead tr:first-child td{background:#4a7696;padding:15px 0 15px}.fluke-pressure-calib-select thead tr:first-child td:nth-child(1n+2) img {max-width:68px}.fluke-pressure-calib-select thead tr:first-child td:first-child{background:#fff}.fluke-pressure-calib-select tbody tr td:first-child{background:#95abc3;text-align:left}.fluke-pressure-calib-select tbody tr:first-child td{background:#d0dbe5;color:#416a86;font-weight:bold}.fluke-pressure-calib-select tbody td:nth-child(even) {background:#d0dbe5}.fluke-pressure-calib-select tbody td:nth-child(odd) {background:#e6ecf2} Application Selection Guide Model number 754 721/?721Ex 719Pro 719 718 717 700G 3130 2700G Dead-weight Testers Application Calibrating pressure transmitters (field) ? ? Ideal ? ? ? ? Calibrating pressure transmitters (bench) ? ? ? ? ? ? Ideal ? Calibrating HART Smart transmitters Ideal Documenting pressure transmitter calibrations Ideal Testing pressure switches in the field Ideal ? ? ? ? ? ? Testing pressure switches on the bench ? ? ? ? ? ? Ideal Documenting pressure switch tests Ideal Testing pressure switches with live (voltage) contacts Ideal Gas custody transfer computer tests ? Ideal ? Verifying process pressure gauges (field) Ideal ? ? ? ? ? ? Verifying process pressure gauges (bench) ? ? ? ? ? ? ? ? Ideal Logging pressure measurements ? Ideal ? Testing pressure devices using a reference gauge Ideal Hydrostatic vessel testing Ideal Leak testing (pressure measurement logging) ? Ideal Pressure Calibration Selection Guide

Downloads

718_30G.pdf_18.pdf

pdf

Specifications

Range
-12 to 30 PSI, -850 mbar to 2 bar (internal sensor) Over Pressure 3xFS
Resolution
0.001 psi, 0.1 mbar
Accuracy
0.05% full scale
Functions
Zero, Min, Max, Hold, Damp, Note: for use with non-corrosive gasses
Loop Power Range
24 V dc
Loop Power Accuracy
+/- 10%, Note: Drive; 20 mA into 1000 Ω for battery >6.8V; 700 Ω for battery 5.8 to 6.8 V
Measure Accuracy Range
0 to 24 mA
Measure Accuracy Resolution
0.001 mA
Measure Accuracy
0.025% + 1 count
Display
LCD, 5 digit pressure and current simultaneous
Operating Temperature
-10 to 55°C
Operating Altitude
3,000m max
Safety
CSA C22.2 No. 1010.1:1992
Maximum Voltage
30V
Size
216 mm L x 94 mm W x 67 mm (8.5 in. L X 3.72 in. W X 2.61in.)
Weight
Approx. 992g (35 oz)
Shock & Vibration
Random, 2G, 5-500 Hz, 1m drop test
Battery Type
Two 9V batteries, ANSI/NEDA 1604A or IEC 6LR61, Battery life: 4 to 20 hours, typical, depending on functions